Skip to content
ScatterwaysDTS / DAS interrogator
Technical details

How it works, and the numbers.

What it measures, how it works, what comes out of the box, what it needs in the field, and the full specification.

01

What it measures

Temperature (DTS). Raman backscatter, anti-Stokes to Stokes ratio, on single-mode or multimode fibre. Home range 0.5 to 5 km on single-mode; multimode to about 10 km at coarser resolution. Spatial resolution in the 0.5 m class, sampled every 0.41 m. Noise 0.11 K at 2 km in five minutes. Single-ended, double-ended loop or mirror-ended layouts through an optical switch. Two Peltier-held reference coils inside the box and double-ended reading through the loop make the calibration automatic on your own cable.

Acoustics (DAS). Coherent-phase Rayleigh backscatter on single-mode fibre: quantitative strain, not only event detection. 40 to 50 km single-ended. Gauge length 1 to 40 m in software, channels from 0.41 m native with programmable stacking, pulse rates to 20 kHz. GNSS pulse-per-second and a 10 MHz reference input for arrays of several units.

Not this instrument. Brillouin strain. Fibre Bragg gratings. Fibres carrying live telecom traffic, where the pump pulse is too strong for a shared fibre. Fire-detection certification. Acoustic raw data comes as decimated streams and raw snapshots as standard; continuous raw streaming is an option.

02

How it works

  1. 1

    A pulse of light goes in

    The instrument sends short pulses of laser light into your fibre. Any single-mode or multimode cable with a common APC connector will do.

  2. 2

    Every point along the fibre answers

    A tiny fraction of the light scatters back from every point along the glass. The arrival time says where it came from. The colour balance of the returning light gives the temperature at that point.

  3. 3

    You get a profile, every half metre

    Many pulses are averaged into one temperature profile along the whole cable. With the acoustic tile, the phase of the returning light is read as well, which turns the same fibre into thousands of strain sensors.

Calibration is automatic on your own cable. Two temperature-controlled reference coils inside the instrument give the absolute scale, and reading a looped cable from both ends removes the cable’s own loss. The method agrees with the open-source dtscalibration reference to 0.014 K on the same raw data, and the files it writes load into that toolchain without conversion.

Temperature

  1. A short pulse of light goes into the fibre.
  2. Every point along the glass sends back a faint glow, in two colours.
  3. The balance of the two colours is the temperature at that point.
  4. The time the glow takes to come back says where it came from.

Acoustics

  1. The same kind of pulse goes in, thousands of times a second.
  2. The light that comes back carries a phase, like the hand of a clock, for every half metre.
  3. When the ground moves, the glass stretches by a few parts per billion and the hand turns.
  4. Every half metre becomes a vibration sensor; plotted over time, events draw themselves.
03

Resolution and noise

Resolution and noise trade against each other, and against averaging time. The modes are a software setting, so you can switch between them from one profile to the next, or run a fine profile once an hour and a low-noise one in between. Averaging longer lowers the noise further: four times the averaging halves it.

Multimode fibre reaches about 10 km at coarser resolution. On single-mode fibre the home range is 0.5 to 5 km. Every rental starts with an acceptance measurement on your own cable, so the numbers you plan with are the numbers you get.

ModeResolutionNoise at 2 kmUse it for
Fine0.43 m0.154 Kthin layers, short hot spots, borehole interfaces
Standard0.48 m0.109 Kmost campaigns
Low noise0.61 m0.077 Ksmall temperature differences over long reaches

Table 1. Temperature modes on 2 km of single-mode fibre, 300 s averaging. Resolution is the 10 to 90 % response to a step; noise is the standard deviation at the far end.

04

Three configurations, one platform

DTSDASDTS + DAS
MeasuresDistributed temperatureQuantitative acoustic strainBoth, time-shared on one fibre
FibreSingle-mode and multimodeSingle-modeSingle-mode and multimode for temperature; single-mode for acoustics
Fibres per instrument2 standard; 4 or 8 with the switch option14 standard
Typical useHydrology, glaciology, geothermalDark-fibre seismology, research arraysWells, pipelines, combined studies

The acoustic capability is one insert tile. A temperature unit becomes a combined unit on site by adding the tile; nothing goes back. Enclosure, software and data formats are the same in every configuration.

05

What comes out: dashboard, API, files

dimensions:  x (bins, 0.41 m)        time (profiles)
variables:
  st, ast            Stokes, anti-Stokes             counts   (x, time)
  rst, rast          reverse-direction channels      counts   (x, time)   double-ended only
  tmp                calibrated temperature          degC     (x, time)
  tmp_var            per-bin variance                K^2      (x, time)
  ref_coil_1, _2     Peltier-held reference coils    degC     (time)
  ext_pt100_1..4     external RTD channels           degC     (time)
  acq_time           averaging time per profile      s        (time)
attributes:  instrument serial, firmware version, acquisition settings,
             fibre preset, switch channel, calibration method, GNSS fix

Listing 1. Variables in the netCDF file written per measurement. The layout follows the dtscalibration convention, so the file loads without conversion.

netCDFdtscalibration-compatible; one file per channel per period
CSVdistance, temperature, variance per profile
HDF5temperature and acoustic containers with the same metadata
miniSEED, SEG-Yacoustic channels for seismology toolchains
PRODML, WITSML-shapedwell-monitoring systems
Modbus TCP, MQTT, RESTlive values and alarms to plant systems

Every rental includes a web dashboard with live profiles, time series, alarms and the campaign report, and an API. REST for profiles, events and status, MQTT for live values, Modbus TCP for control rooms. The same data is written as files in the formats listed.

The instrument runs in one of four modes: pushing to a cloud account, serving on the site network, writing to a local disk only, or air-gapped with configuration by USB stick. The measurement chain is identical in all four; the cloud is a sink, never a dependency. When a link drops the unit keeps its schedule and catches up afterwards.

06

Power in the field

ModeAverage drawOn 100 AhNotes
Temperature, one 5-min profile per hour2.3 Wabout 20 daysa 150 W panel carries it February to October at Danish latitude
Temperature, one profile per 4 h0.8 Wabout 2 monthsglacier and permafrost schedules
Temperature, continuous21 Wabout 2 daysuse PoE or mains
Acoustics, continuous25 Wunder 2 dayscannot duty-cycle; mains, fuel cell or a large solar kit

Table 2. Average draw and endurance on a 12 V, 100 Ah lithium iron phosphate battery (about 1.1 kWh usable).

Inputs are 9 to 36 V DC and 802.3bt power over Ethernet. Between profiles the instrument sleeps and only the scheduler and the reference coils stay powered. If the supply fails, the files are closed cleanly and the schedule resumes when power returns.

07

Specifications

Temperature mode
Range0.5 to 5 km single-mode; multimode to about 10 km at reduced resolution
Spatial resolution0.5 m class; 0.48 m 10-90 % on 2 km single-mode
Noise0.11 K at 2 km, 300 s averaging
Sampling0.41 m
Acceptance benchmark0.20 K at 0.6 m, 2 km single-mode, 300 s; or 0.20 K at 1.0 m, 600 m multimode, 60 s
Updateseconds (coarse) to minutes (precise)
Fibresingle-mode G.652 and G.657; multimode 50/125 and 62.5/125
Layoutssingle-ended, double-ended loop, mirror-ended; switch 1x2, 1x4, 1x8
Acoustic mode
Methodcoherent phase, quantitative strain
Range40 to 50 km single-ended
Gauge length1 to 40 m, software selectable
Channel spacing0.41 m native; programmable stacking
Pulse rateto 20 kHz, range dependent
SynchronisationGNSS PPS in and out, 10 MHz reference in
Optics
Wavelength1550 nm
Laser safetyClass 3B; hardware interlock; automatic power reduction on fibre break
Connectorsswappable bulkhead adapters FC/APC, SC/APC, E2000/APC, LC/APC
Power and environment
Inputs9 to 36 V DC; 802.3bt PoE
Draw2.3 W averaged at one profile per hour; 21 W continuous temperature; 25 W continuous acoustics
EnclosureIP67 field case, fanless; 19-inch rack variant on request
Ambient-20 to +50 C
Data and interfaces
Network1 GbE; LTE modem option; Wi-Fi for commissioning
FilesnetCDF (dtscalibration), CSV, HDF5, miniSEED, SEG-Y, PRODML and WITSML-shaped
LiveModbus TCP, MQTT, REST
Modescloud, LAN, disk-only, air-gapped; store-and-forward
Field I/O2x RS-485, SDI-12, 4x 4-20 mA, 4 external and 2 internal PT100, 4 pulse inputs, 4 relays
Raw accessdecimated streams and raw snapshots standard; continuous raw streaming optional

Outside these numbers? Ask.

Longer fibre, finer resolution, a faster update, an unusual connector, a sensor or a data path we have not listed, a site with no power at all. Much of what looks out of spec is a configuration, a tile or a cable away, and we build to order where it is not. Tell us the question and the site. If we cannot do it, we will say so and point you to who can.

Ask about a bespoke setup
08

Questions

Can I use my existing cable?
Usually. Single-mode G.652 and G.657, multimode 50/125 and 62.5/125, any of the common APC connectors. Tell us the fibre type, length, connectors and how the ends are reached; for acoustics the cable's coupling to the ground matters more than the fibre. No cable on site? We supply armoured, borehole and hybrid single-mode plus multimode sensing cable, terminated and cut to length, with the instrument.
How is it calibrated?
Two Peltier-held reference coils inside the instrument give the absolute scale; reading the loop from both ends removes the cable's differential loss; external PT100 inputs let you add calibration sections of your own. No baths in the field.
What if it does not perform on my fibre?
Every rental starts with an acceptance benchmark run on your own cable: 0.20 K at 0.6 m over 2 km of single-mode fibre with five-minute averaging, or 0.20 K at 1.0 m over 600 m of multimode with one-minute averaging. You see the result before the campaign starts, and every figure we report carries the fibre, the length and the averaging time it was taken with.
Does it run on battery?
Yes, for temperature. At one profile an hour it draws about 2.3 W and lasts about 20 days on a 100 Ah battery; a 150 W panel keeps it going most of the year at Danish latitude. Continuous acoustics needs mains, PoE or a fuel cell.
Can I get raw data?
Decimated acoustic streams and raw snapshots as standard. Continuous raw streaming is an option; tell us the data rate your processing needs.
Can I buy one outright?
No. Scatterways is rented by the month: the instrument, the cable if you need one, installation help, remote operation and processing, in one monthly line. Campaigns from two weeks, seasons month by month, continuous monitoring on a service contract.
Do you install on site?
If you want us to, yes. The default is simpler: the box ships, your technician or student connects it on a video call with us, and we operate it remotely. Air-gapped sites get a local operating plan instead.
Can several units run as one array?
Yes. Each unit takes GNSS pulse-per-second and a 10 MHz reference, so units on separate fibres share one clock and their acoustic channels line up to the sample.
Is it safe?
The laser is Class 3B behind a hardware interlock. Power drops automatically if the fibre breaks or a connector is opened, and the instrument reports it.

Put your fibre to work.

Tell us about your fibre and your question. You get a call, a plan for your site and a monthly price.